Stability test chamber

Stability test chamber

Stability test chamber

FG Stability test chamber ZF model is the specialist for reliable stability testing and precise maintenance of constant climatic Conditions. From programming to documentation, it meets all important guidelines, such as ICH, GMP, and GLP. With its high capacity reserves, automatic water and wastewater management, and many options, the FG ZF series is well-prepared for the future.

sms alarm

This device is used in various industrial pharmaceutical research laboratories

HCT (Homogenic chamber technology)

Temperature and humidity stability and reproducibility
Uniform circulation even under full load
Short warm-up time
Short recovery time
Triple-walled

DUAL SAFETY

Class 3.3 independent temperature safety device (DIN12880) (TWW)
Class 3.1 independent temperature safety device (DIN12880) (TWB)
Audible and Visual alarm

NON-CUT technology

Powerful cooling system for safe operation up 30°C ambient temperature
Dual evaporator
Frost-free

Flexible water management with anti-polar technology

Water sensor protection thanks to anti-polar technology
Electric sensor for adjust water surface
Prevent water deposition on sensors

Temperature and humidity chamber humidifier system

Unique humidifier system

Vapor pressure humidification with fast response time
Solution independent of installation sight
Pump for discharges up to 1 m in height
Short recovery times after door opening
Drift-free, capacitive humidity sensor
Finely adjustable humidity control
External water supply

ARM series micro controller

Save temperature, humidity, fan speed and all events up to 5 years
Controller which can store 3 programs of 100 steps
6 points user temperature calibration
Double PT100 temperature sensor
2 points user humidity calibration
PID controller with high accuracy
7”Color graphic TFT LCD
Capacitive touch key
Lock functions

STM (smart temperature monitor)

Alarm for temperature and humidity
Audible and visual alarm
Door open alarm
SMS alarm

DOCUMENTION

Internal memory to record the values of temperature, humidity and fan speed for up to 2 years
Record values according to GMP and GLP standards
Draw a graph of temperature and humidity on the device's LCD

FGLAB software

The ability to print graphs in a specific time frame in accordance with GMP and GLP standards, without the need to install intermediate software, the ability to connect the device to a computer, the ability to view values online,

Product video

To get to know More details of the product Watch the opposite video.

Product video gallery

ZF model temperature and humidity test chamber technical specification table

External specifications
ZF120
ZF240
ZF400
ZF700
ZF1400
Depth with a handle and outlet water source (cm)
95
95
120
129
Depth with handle (cm)
75
75
75
90
112
Height of the device with the base (cm)
106
148
204
197
205
Device width (cm)
82.5
90
90
121
160
number of doors
1
1
1
2
2
Internal specifications
ZF120
ZF240
ZF400
ZF700
ZF1400
Internal volume (Lit)
120
400
700
1342
width(cm)
59
66
66
97
136
depth (cm)
46
45.5
45.5
57
77
Empty weight of the device (Lit)
115
150
220
450
700
Weight capacity for each tray (kg)
5
5
5
5
10
Ability to bear the weight of the entire floor (kg)
25
25
35
30
30
The number of standard trays in the machine - (can be ordered)
2-(2)
2-(6)
2-(13)
2-(12)
2-(5)
Temperature and humidity specifications (*)
ZF120
ZF240
ZF400
ZF700
ZF1400
Temperature recovery time after 30 seconds of opening the door at 25°C and 60% humidity (min)
10
10
20
20
Temperature recovery time after 30 seconds of opening the door at 25°C and 60% humidity (min)
±1
±1
±1
±1
±1
Uniformity of temperature with humidity at 25°C (Variation) (±K)
±1
±1
±1
±1
±1
Fluctuation of temperature without humidity at 25°C (fluctuation) (±K)
±0.5
±0.5
±0.7
±0.5
±0.5
Temperature fluctuation with humidity at 25°C (fluctuation) (±K)
±1
±1
±1
±1
±1
Temperature fluctuation without humidity at 40°C (fluctuation) (±K)
±0.6
±0.6
±0.6
±0.6
±0.6
Temperature fluctuation with humidity at 40°C (fluctuation) (±K)
±0.6
±0.8
±1
±1
±1
Temperature uniformity without humidity at 40°C (Variation) (±K)
±1.5
±1.5
±1
±1.5
±1.5
Uniformity of temperature with humidity at 40°C (Variation) (±K)
±1.5
±1.5
±1.5
±1.5
±1.5
Temperature range without humidity (°C)
0-60
0-60
0-60
0-60
0-60
Uniformity of temperature with humidity at 40°C (Variation) (±K)
±1.5
±1.5
±1.5
±1.5
±1.5
Temperature range with humidity (°C)
10-60
10-60
10-60
10-60
10-60
humidity fluctuation (%RH)
±5
±5
±5
±5
±5
Humidity Range (%RH)
30-90
30-90
30-90
30-90
30-90
Installation conditions
ZF120
ZF240
ZF400
ZF700
ZF1400
Distance of the device from devices and side walls (cm)
15
15
15
15
Free space above the device (cm)
30
40
40
40
40
The distance of the device from the back wall (cm)
20
20
20
20
20
Minimum and maximum ambient temperature (°C)
+16 to +26
+16 to +26
+16 to + 26
+16 to +26
+16 to +26
Maximum humidity (%RH)
70
70
70
70
70
Electronic information
ZF120
ZF240
ZF400
ZF700
ZF1400
Power (W)
1540
3300
3750
4000
Frequency (HZ)
50
50
50
50
50
Voltage(V AC)
220
220
220
220
220
Current(A)
7
12
15
17
18.2

Customized equipment and services

Access port with silicone plug
Qualification folder IQ/OQ/PQ
Additional stainless steel rack
Calibration certificate
Door lock

Drug stability test chambers or temperature and humidity chambers (CLIMATE CHAMBER) are important for creating special environmental conditions for pharmaceutical ICH tests. Drug sustainability tests are an important part of the development of pharmaceutical and medical sciences, doubling the chambers’ importance and sensitivity. Stability chambers are mainly used for ICH GUIDELINE Q1A tests, which place the drug in temperature conditions of 40 degrees Celsius and 70% humidity, and measure the stability of the drug in 3-month, 6-month, and even 1-year periods.

The designers of this series of chambers have tried to meet the needs of a wide range of users along with compliance with international standards. For this purpose, facilities such as system DUAL SAFETY (safety thermostat)، Unique HCT technology (Uniformity in different parts of the chamber)، ARM series advanced microcontroller، NON-CUT technology (for accuracy and temperature stability in cooling mode)، ANTIPOLAR technology (to prevent damage to water level sensors) In this series of temperature and humidity chambers are placed. Also, compliance with ICHGMP, and GLP standards is among the other advantages of these chambers.

The FG design manager assures you that this drug stability test chamber series is fully competitive with European samples, which you will find after reviewing this series of chambers.

In general, environmental condition test chambers or drug test chambers are divided into different types, the smallest volume of which is desktop to the largest volume of which is from about 5000 liters to 25000 liters, and they are called WALK-IN chambers.

As mentioned before, CLIMATE CHAMBERS are part of the important equipment of laboratories and production lines and the quality control department of pharmaceutical factories. But in addition, the chambers are also used in the laboratories of food, health, and cosmetic industries.
However, due to the simultaneous supply of temperature and humidity, they are also used in university laboratories and research and industrial centers. The external body of the chamber is made of iron sheet with electrostatic powder paint, and the inner chamber is entirely made of stainless steel. In addition to its extraordinary strength, it has given the device a special beauty. By carefully looking at the inner compartment of the device and observing different angles, we notice the use of curved corners inside the device, which makes it easy to clean and safe for the user.
Also, using stainless steel makes the inner chamber resistant to humidity. Conversely, the ultra-advanced device controller comes with the latest ARM series microprocessors. The device’s controller significantly contributes to temperature control and stability of temperature and humidity by using smart algorithms.

The temperature and humidity chambers must have a precise temperature and humidity controller, a circulating fan for the uniformity of temperature and humidity inside the chamber, the possibility of warning in case of emergency, the ability to program, the ability to adjust the thermal gradient, the screen to show information, the ability to set the timer. And to be able to save information.

In some cases, it has been observed that the deviation of the temperature and humidity inside the chamber from the set values ​​causes damage to the test samples and sometimes a lot of damage, so the device must be equipped with a separate system to prevent the unwanted increase or decrease of temperature and humidity and to notify It is to the user in case of occurrence.

Another important point of the stability test chamber is the ability to calibrate the device. If the calibration is 1 point, the device is calibrated only at the same point, and by calibrating at one temperature, the calibration of different temperatures is out of adjustment.

Some devices are calibrated in two temperatures; in this case, the device considers a hypothetical line based on these two temperatures and calibrates the rest of the temperatures based on this hypothetical line. Since some points may not be in line with the hypothetical line of the device, calibration at two points cannot be guaranteed for the high accuracy of the device, so the higher the calibration points, the more accurate the calibration is done and the device works in all precise temperature range.

Another point is that you must have encountered the phenomenon of an unwanted increase in chamber temperature and failure of laboratory samples. Today, international standards emphasize the control of the temperature of the chamber by an independent thermostat so that when necessary, by cutting the voltage of the element, they prevent the increase of the temperature of the chamber and prevent possible damage. The ability to connect to FGLAB software is another advantage of this device.

By connecting to this software, you can view all device information, including temperature, humidity, and fan speed, in the form of graphs and tables. Zooming in a specific time range and printing are other advantages of these chambers.

All FG stability chambers have fans and are produced in 4 models of 120, 240, 400, and 700 liters in terms of volume, which, upon request, can also be made according to customer needs. Also, according to the requirements of pharmaceutical factories and the needs of respected customers, it is possible to provide QUALIFICATION documents, including IQ, OQ, and PQ.

All FG stability chambers have fans and are produced in 4 models of ZF120E, ZF240E, ZF400E, ZF700E, and ZF1400E liters in terms of volume, which can also be made according to customer needs upon request. Also, according to the requirements of pharmaceutical factories and the needs of respected customers, it is possible to provide QUALIFICATION documents, including IQ, OQ, and PQ.

We have listed for you what separates our sustainability chamber from other brands:

  • Beautiful and ergonomic design
  • Circulating fan with adjustable speed
  • Easy calibration of humidity in 2 points and temperature in 6 points by the user
  • Controller lock to prevent accidental changes
  • PID controller to control temperature and humidity
  • Controller with 3 programs of 100 steps
  • Separate independent safety sensors to ensure the device’s temperature (with two PT100 sensors).
  • A separate mechanical thermostat according to DIN12880 class 3,3 standard (TWW)
  • Connect the device to the computer through the LAN port
  • No need to install the driver
  • The ability to connect to a computer via WIFI
  • The ability to work with FGLAB software through mobile phones, tablets, and computers
  • Internal memory to record the values ​​of temperature, humidity, and fan speed for up to 2 years
  • Record values ​​according to GMP and GLP standards
  • Drawing a temperature and humidity graph on the device’s LCD screen

ZF 120 E, ZF 240 E, ZF 400 E, ZF 700E, ZF 1400 E

Frequently Asked Questions(FAQ)

Considering that the water source of the stability chambers is separate from the device, it is possible to disconnect the source from the device when entering the laboratory and reconnect it after installing the device, in which case the depth of the device without the source will be equal to 75 cm.

According to FG standards and the world’s standards, the best distance from the walls, floor, and ceiling of the stability test chamber is 10% of the length of the same side; for example, if the width of the device is 40 cm, 10% is 4 cm from each side (left and right). ) and if the depth of the machine is 50 cm, 10% of that means 5 cm from the front and back, the distance from the walls of the machine should be considered for loading.

It always takes time for the internal chamber temperature to reach the set point. The Smart timer is activated when the temperature reaches the set point.

Since many centers do not have facilities such as water or sewage pipes, a feature has been placed in the FG stability chambers so that they can be installed and work properly independently of the deficiencies of the environment. Therefore, the FG stability test chamber is equipped with an inlet water source. And it is an exit and does not need plumbing.

Alarms should be such as to attract the user’s attention. Now imagine a device that repeatedly and unnecessarily alarms, such as when the door is opened (naturally, the device’s temperature goes higher or lower). In normal devices, the alarm is activated, but in the FG device, the alarm is activated only when needed, and in The malfunctioning device is activated.

No, using HCT technology, moist air enters the device instead of water vapor entering the chamber, preventing the samples from getting wet.

Yes, it can be done by the consumers and according to the request.

Relative humidity has a direct relationship with heat, so there are always restrictions on devices that control temperatures and humidity. FG devices are no exception, so it is recommended to chart the humidity temperature in the database of the devices FG is referred to.